Model Number MI2355A |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during implant, there was suspected blockage of the introducer which resulted in a failure to aspirate.The physician tried to withdraw the sheath a few centimeters and re-attempt after putting back the dilator into the sheath.Still not able to aspirate.The introducer was attempted not used and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Product event summary: the introducer was returned and analyzed.The distal seal of the delivery system introducer was torn.The analyst noted the introducer was returned with the dilator separated from the introducer.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|