• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G447
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Event Description
It was reported that noise oversensing on the right ventricular (rv) lead, resulting in inappropriate anti-tachycardia pacing (atp).Provocation maneuvers were performed, and all measurements remained within normal limits, with no noise present.The patient reported they put their cellphone in the shirt chest pocket, which is suspected to be electromagnetic interference (emi) and the source of the noise.The patient was instructed to keep a distance between the crt-d and cellphone, and further remote monitoring is expected.This crt-d remains in-service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19071284
MDR Text Key339717480
Report Number2124215-2024-21488
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/26/2022
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number235021
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-