Model Number 0600-0783
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2024
Event Type  malfunction  
Manufacturer Narrative
A replacement glidescope core smart cable was provided to the customer and the reported glidescope core smart cable used during the reported event was returned to verathon for evaluation.A verathon technical service representative evaluated the returned smart cable and was able to confirm the reported cable failure.The smart cable passed visual inspection.However, when connected to known, good, test verathon equipment, the smart cable produced poor quality image (pixelated/discolored image).Manipulating the connection between the smart cable's hdmi connector and the imaging device caused the image to split.The customer's glidescope core smart cable failed verathon's device functionality testing.Upon completion of verathon's device evaluation, the customer's smart cable was scrapped due to already being provided a replacement and there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope core smart cable, there was intermittent loss of image while intubating the patient.No delay in procedure, use of a back-up device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas rd
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas rd
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n. creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key19073085
MDR Text Key340588627
Report Number9615393-2024-00057
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-