COVIDIEN MFG DC BOULDER VALLEYLAB FT10; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number VLFT10GEN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Superficial (First Degree) Burn (2685); Partial thickness (Second Degree) Burn (2694)
Event Date 03/09/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
According to the reporter, at the beginning of the drill hole for chronic subdural hematoma under local anesthesia, after a couple of activations, they saw a flame from under the surgical drape which set the hair on fire resulting in a 2nd degree burn to the face, ears, and neck, the second degree burn site on the neck measured approximately 20cmx5cm, the first degree burn site on chin measured approximately 5cmx5cm.The flame was about 20cm high on the hair and was extinguished with water.The generator was replaced to complete the procedure and the burn was treated with medication, future plastic surgery to repair the damage should be evaluated.Patient hospitalization was extended for 10 days.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB FT10
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19074075
MDR Text Key339773030
Report Number1717344-2024-00932
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10884521516328
UDI-Public10884521516328
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K151649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVLFT10GEN
Device Catalogue NumberVLFT10GEN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/09/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight62 KG
-
-