SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: unknown.No information has been provided to date.Initial reporter occupation: field service ireland & uk: service investigation summary: the affected device was returned for evaluation and visual inspection was performed.The device had broken leds, and 3 rubber feet were missing, confirming the customer's indicated failure.Functional and electrical safety tests as per qcp 026 were performed - device failed alarm check.The most probable root cause of this problem is mishandling.As a result, the pwa was replaced.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
|
|
Event Description
|
It was reported that during preventative maintenance, the bottom 2 leds were found broken from the main board.There was no patient involvement and no harm/adverse event reported.
|
|