Model Number HL-90
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.No information has been provided to date.Initial reporter occupation: field service ireland & uk: service investigation summary: the affected device was returned for evaluation and visual inspection was performed.The device had broken leds, and 3 rubber feet were missing, confirming the customer's indicated failure.Functional and electrical safety tests as per qcp 026 were performed - device failed alarm check.The most probable root cause of this problem is mishandling.As a result, the pwa was replaced.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
 
Event Description
It was reported that during preventative maintenance, the bottom 2 leds were found broken from the main board.There was no patient involvement and no harm/adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key19075617
MDR Text Key339776542
Report Number3012307300-2024-02306
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)150908
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90-UK-230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2024
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-