Model Number HL-90
Device Problems Fracture (1260); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
Date of event is unknown; no information has been provided to date.One device was received.Per visual inspection, damaged connection between the power switch and the printed circuit board (pcb), cracked front cover, 2 standoffs behind the liquid crystal display (lcd) were broken, light emitting diode (led) lights were broken, non-oem power cord, outdated quick connect, pcb, and power switch.Functional testing confirmed the complaint as the device has no power.The root cause was a damaged pcb.It was unknown what caused it to become damaged.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Event Description
It was reported that the unit would not turn on.When they looked at the main panel it was cracked and broken.Some parts on the main board are potentially damaged.It was unknown when the fault occurred no patient was involved, and no patient was harmed as the unit was not able to power on.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19075884
MDR Text Key339776661
Report Number3012307300-2024-02308
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2024
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-