Date of event is unknown; no information has been provided to date.One device was received.Per visual inspection, damaged connection between the power switch and the printed circuit board (pcb), cracked front cover, 2 standoffs behind the liquid crystal display (lcd) were broken, light emitting diode (led) lights were broken, non-oem power cord, outdated quick connect, pcb, and power switch.Functional testing confirmed the complaint as the device has no power.The root cause was a damaged pcb.It was unknown what caused it to become damaged.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
|