The customer reported a replace sensor error message.Abbott diabetes care (adc) attempted to replicate the reported issue and the reported configuration of android 14 is not compatible with the freestyle librelink application.The compatibility guide is available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer, and the incompatible configurations were used, this complaint is therefore not confirmed to use.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
An error message was reported with the adc device in use with samsung a23 phone with android operating system version 14.A customer reported a "sensor error" error message with the adc device and was unable to obtain readings.The customer was provided juice, sugar, and insulin (dose/type unknown) by a non-healthcare professional.There was no report of death or permanent injury associated with this event.
|