A low reading issue was reported with the adc device.The customer's caregiver reported that they received unspecified lower sensor scan results compared to unspecified readings obtained on a healthcare meter.As a result, the customer experienced a loss of consciousness.The customer had contact with a healthcare professional (hcp) who obtained a reading of (278 mg/dl) on a healthcare meter.There was no report of any third-party or hcp treatment.There was no report of death or permanent injury associated with this event.A sensor scan result of 43 mg/dl was obtained and compared to a healthcare meter result of 278 mg/dl.The results when plotted on a parkes error grid fall into the "d" zone, showing the difference in values to be clinically significant.It is unknown when these readings were obtained in relation to the reported adverse event.
|
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the libre sensor and sensor kit were reviewed, and the dhrs showed the libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|