Catalog Number UNK SHOULDER HUMERAL CUP |
Device Problem
Naturally Worn (2988)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/02/2024 |
Event Type
Injury
|
Event Description
|
Patient revised to address humeral cup polyethylene wear.During the removal of an angle stable screen from the holding plate, the teeth of the screwdriver broke off.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.Additional event information: after the screwdriver broke, the well-holding metallene remained in position and the glenosphere and inlay have been changed.The inlay was worn on one side.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed, as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected and distributed to approved specifications.Additional complaint information, monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown.Therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|
|
Search Alerts/Recalls
|
|