MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE Back to Search Results
Catalog Number 50000000E
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/02/2024
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that nurse stated they started cooling yesterday at 19:30.The device was alarming 77/78 (non-recoverable system error).Chiller water temperature sensor out of range.Instructed nurse to reboot the device.Nurse stated they did reboot it once already and the alarm returned, they are currently rebooting it now.Device powered back on, non-recoverable alarm returned.Informed nurse they would have to swap to another device and send this one to biomed for evaluation.Nurse stated they have 66 hours remaining in cooling.Informed nurse once they select hypothermia on the new device, they could adjust the cooling settings and change the duration to 66 hours.Informed nurse because they cannot empty the pad, they could disconnect it over a trashcan to collect the water.Confirmed with nurse they could use the same pad.Per sample evaluation results received on (b)(6) 2024, it was reported that hot side connection between the power inlet module and the ac main voltage circuit card shows signs of electrical overstress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19078674
MDR Text Key340601967
Report Number1018233-2024-01962
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-