• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G158
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2024
Event Type  malfunction  
Event Description
It was reported that this patient was scheduled to undergo a non conditional magnetic resonance imaging (mri) procedure.This system has non boston scientific leads.Upon programming into mri protection mode, it was noted that the left ventricular (lv) lead exhibited high out of range pace impedance measurements.No adverse patient effects were reported.At this time, this device system remains in service.
 
Event Description
It was reported that this patient was scheduled to undergo a non conditional magnetic resonance imaging (mri) procedure.This system has non boston scientific leads.Upon programming into mri protection mode, it was noted that the left ventricular (lv) lead exhibited high out of range pace impedance measurements.Further information was received that the patient did not receive the mri.There was a device check after the initial mri was canceled and out of range measurements were again obtained.No adverse patient effects were reported.At this time, this device system remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19079893
MDR Text Key339907722
Report Number2124215-2024-21799
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG158
Device Catalogue NumberG158
Device Lot Number322453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-