Catalog Number L-70
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Event Description
It was reported that water was leaking from the tip of the patient connection.This occurred during priming, no patient involvement and no patient harm/adverse event reported.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Additional information: h4 - manufacture date.H3 and h6 - evaluation codes: updated.Device evaluation: one (1) unit received with original package.Two photos were also submitted.Picture one shows an l-70 device, picture two shows a return connector.In the inspected sample, delamination is observed at the junction of the return connector and the tube which could cause leakage.The unit failed the leak test.The reported failure mode is confirmed.Based on the analysis on the returned unit, and review of the mitigations during the manufacturing process, the root cause of the lack of solvent.Awareness notification for failure mode reported to production personnel was issued.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW HOTLINE DISPOSABLE SET
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19082442
MDR Text Key340139939
Report Number9617604-2024-00302
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier30695085407007
UDI-Public30695085407007
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberL-70
Device Lot Number4293567
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-