Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that on (b)(6) 2024, during the procedure, the surgeon was attempting to ream the medullary canal of the femur when the 10.5mm medullary reamer head became disconnected from the flexible shaft within the canal.The surgeon was not using a reaming rod with ball tip which the surgical technique states to never do.The surgeon was unable to retrieve the reamer head within the canal.The procedure was successfully completed.This report is for one (1) 5.0mm flexible shaft this is report 2 of 2 for complaint (b)(4).
|