MEDTRONIC EUROPE SARL COBALT¿ XT HF QUAD CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTPA2QQ |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Erosion (1750); Erythema (1840); Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
|
Event Date 03/20/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: 6935m55 lead implant date: (b)(6) 2018 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
It was reported that the patient experienced an infection, erosion, pain, redness, and swelling.The cardiac resynchronization therapy defibrillator (crt-d) system was explanted and replaced six days later. no further patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|