• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG UNIPOLAR HIGH FREQUENCY CORD, 300 CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG UNIPOLAR HIGH FREQUENCY CORD, 300 CM Back to Search Results
Model Number 26006M
Device Problems Fire (1245); Flare or Flash (2942)
Patient Problem Burn(s) (1757)
Event Date 03/12/2024
Event Type  Injury  
Manufacturer Narrative
Attempts were made to obtain more information and to have reported device return for evaluation; however, the reporter has not responded to the requests at the time of this report.If the reported device will be returned or the evaluation is completed, a supplemental report would be made to the fda.This complaint will be tracked and trended.The event is filed under internal karl storz complaint id: (b)(4).
 
Event Description
It was reported that the unipolar high frequency cord sparked and caught fire during the surgery.While using the electrode to cauterize, the center area of the cord sparked and caught surgical drape and arm cradle foam padding on fire.A small amount of water was used to extinguish the fire.The patient sustained a burn on he inner left forearm and also several spots on the left side of abdomen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIPOLAR HIGH FREQUENCY CORD, 300 CM
Type of Device
UNIPOLAR HIGH FREQUENCY CORD, 300 CM
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave.
el segundo, CA 90245
MDR Report Key19085265
MDR Text Key339875515
Report Number9610617-2024-00097
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K945380
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26006M
Device Catalogue Number26006M
Device Lot NumberWZ01
Is the Reporter a Health Professional? No
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR
Patient SexMale
Patient RaceWhite
-
-