Catalog Number 2B5LT
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date sent: 4/11/2024.D4: batch # a9d717.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the 2b5lt device was received with the sleeve broken.The piece broken from the sleeve was returned.In addition, the packaging opened was returned along with the instrument.No conclusion could be reached regarding what may have caused the reported incident.One possible cause for the damage found may be due to excessive force being applied to the device.Please refer to the instructions for use for additional information regarding proper device usage.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Event Description
It was reported that during a lap gastrectomy surgery, after insert into the trocar, noted the device was broken.Another device was used to continue the surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT BLADELESS 5X100 STABILITY
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key19087539
MDR Text Key340624564
Report Number3005075853-2024-02813
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036013709
UDI-Public10705036013709
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2B5LT
Device Lot Number494C46
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2024
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-