Catalog Number 07976887190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Manufacturer Narrative
The cobas 8000 core unit serial number is (b)(6).The serial number for the customer's e801 module was not provided.
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ft4 iii (ft4 iii) on a cobas 8000 core unit compared to the wako accuraseed and abbott alinity methods.Refer to the attached data for the patient results.
 
Manufacturer Narrative
The sample was submitted for investigation and tested on a cobas e801 analyzer at the investigation site.The ft4 result was discrepant when compared to the wako accuraseed and abbott alinity methods.The sample was sent to an external laboratory and tested by liquid chromatography with tandem mass spectrometry (lc-ms/ms) and the ft4 results were comparable to the roche results.See the attached data for the additional results.The e801 serial number used at the investigation site was (b)(6).The ft4 iv reagent lot number used at the investigation site was 749372 with an expiration date of 31-oct-2024.From the information provided, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19088536
MDR Text Key340106175
Report Number1823260-2024-01119
Device Sequence Number1
Product Code CEC
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2024
Initial Date FDA Received04/11/2024
Supplement Dates Manufacturer Received04/14/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-