Catalog Number 80219 |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/18/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during a recanalization procedure in the occluded right superficial femoral artery via the left femoral artery approach, the coil spring of the catheter was allegedly found to be broken in the middle and was inside the patient.Reportedly, the broken part of the catheter was successfully removed using a catcher.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Manufacturer Narrative
|
The catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex that are cleared in the us.The pro code and 510 k number for the rotarex are identified in d2 and g4.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.(medical device lot number).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a thrombectomy and atherectomy procedure in the occluded right superficial femoral artery via the left femoral artery approach, the coil spring of the catheter was allegedly found to be broken in the middle and was inside the patient.Reportedly, the broken part of the catheter was successfully removed using a catcher.The procedure was completed using another device.There was no reported patient injury.
|
|