It was reported, during the unspecified procedure, the forceps on two different thunderbeat devices did not close properly and the tip has broken off inside the patient.Both devices were used on the same patient.It is unknown if the tip was retrieved.Additional procedural details have been requested but not yet provided.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The subject device was manufactured in december 2023 based on the provided 3-digit lot information "3zk", but the specific date is unknown.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and past investigation results, a likely reason for the falling of the grasping section might be the following: 1.An excessive force in the direction for opening was applied to the grasping section.This caused the grasping section to detach from the shaft.2.Since a force in the twist direction was applied to the detached grasping section, the distal end of the inner pipe broke.As a result, the grasping section fell off.However, the subject device was not returned, and a specific root cause could not be identified.The event can be detected/prevented by following the instructions for use which state: "when inserting the thunderbeat into or withdrawing it from the trocar tube, gently hold the control handle and make sure that the grasping section is closed.If the thunderbeat is inserted or withdrawn with the grasping section open, the probe tip/grasping section may become damaged, or it may become impossible to withdraw it from the trocar." "should any crack, scratch, deformation, split, protrusion, or partial separating be observed on the probe tip, grasping section, tissue pad, shaft, the surface of the transducer, transducer cord, or transducer plug, do not use them and replace the damaged instrument or the transducer with a spare.Using a damaged device may cause burns due to abnormal output or high-frequency (rf bipolar) current leakage or breakage of the probe tip, the tissue pad, and the grasping section." olympus will continue to monitor field performance for this device.
|