This event has been recorded by zimmer biomet under (b)(4).G2, country event occurred: slovenia review of the most recent repair record determined the motor speed was unstable, the reciprocation arm was worn, the unit was out calibration at the 0 reading and the control bar position was not correct.The motor and reciprocation arm were replaced and the unit and position of the control bar were recalibrated and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|