ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2, country event occurred: slovenia review of the most recent repair record determined the motor speed was unstable, the reciprocation arm was worn, the unit was out calibration at the 0 reading and the control bar position was not correct.The motor and reciprocation arm were replaced and the unit and position of the control bar were recalibrated and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during testing the dermatome had no power.There was no patient involvement or impact.Due diligence information complete.After evaluation of the returned device, it was determined that the device had inconsistent speed.No adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key19096525
MDR Text Key340028035
Report Number0001526350-2024-00437
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375994
UDI-Public(01)00889024375994(11)190404(10)64371014
Combination Product (y/n)N
Reporter Country CodeSI
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64371014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-