ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER420
Device Problem Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Event Description
It was reported that during a bariatric procedure, clips were crossing.
 
Manufacturer Narrative
(b)(4).Date sent: 4/12/2024.D4: batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: "1.Could you please clarify if there were any patient consequences? 2.Could you please clarify if was any change in the post-operative care of the patient as a result of the event?" attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 5/21/2024.D4: batch # x7041v.Investigation summary: the product was returned to ethicon  for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er420 device was returned with a clip in the jaws; the clip was removed in order to inspect the jaws and they were found to be misaligned.In an attempt to replicate the reported incident, the device was tested for functionality; during the analysis, the instrument was cycled and it fed and formed ten(10) scissored clips due to the misaligned condition of the jaws; finally the instrument locked out as intended.The event reported was confirmed and it is related to improper use of the device.Possible causes for the condition found may be if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key19096622
MDR Text Key340801442
Report Number3005075853-2024-02857
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberER420
Device Lot NumberX7042G
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/20/2024
Initial Date FDA Received04/12/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-