• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUADRA ASSURA MP ICD; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUADRA ASSURA MP ICD; No Match Back to Search Results
Model Number CD3371-40QC
Device Problems Over-Sensing (1438); Pacing Asynchronously (1441)
Patient Problem Arrhythmia (1721)
Event Date 10/28/2023
Event Type  malfunction  
Event Description
It was reported that pacemaker mediated tachycardia and electromagnetic interference resulting in oversensing was observed on the device.Abbott technical support recommended reprogramming to resolve the event, however, no intervention has been performed at this time to resolve the event.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19096756
MDR Text Key340030167
Report Number2017865-2024-37120
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508407
UDI-Public(01)05414734508407(10)A000119811(17)231231
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model NumberCD3371-40QC
Device Lot NumberA000119811
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-