The conclusions are as follows: the visual inspection did not reveal any abnormality.The analysis confirms that the connection system of the subject pacemaker functioned as specified using a duplicate torque-limiting screwdriver.Normal tightening marks were seen on the ventricular setscrew tip.Those tightening marks are most likely related to the first lead connection during the first intervention.The ventricular setscrew was found completely unscrewed.The associated glue bead which is present to avoid losing the setscrew was found cracked.This complete unscrewing performed to reposition the ventricular lead has most probably led to the impossibly to screw again the setscrew.Indeed, once the setscrew is completed unscrewed (out of its cavity) it is difficult to screw again.Based on the available data, no issue is suspected on the subject pacemaker.
|