• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2024
Event Type  malfunction  
Manufacturer Narrative
H11: g1 phone number (b)(6).
 
Event Description
Philips received a complaint by the customer on the v60 indicating that there was a "proximal pressure line disconnect" alarm that had occurred.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.It was reported that there was a "proximal pressure line disconnect" alarm that had occurred.The reported issue was duplicated during a run-in test.An attempt to resolve the issue by changing the circuit was unsuccessful.Onsite service was then given to the customer and the reported issue was unable to be duplicated during a run-in test.The reported issue was unable to be duplicated.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19098130
MDR Text Key340041980
Report Number2518422-2024-19197
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public(01)00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2024
Date Device Manufactured10/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-