It was reported that a large volume infusor leaked within the sealed overpouch; the blue tip (cap) of the tube was "not connected or not securely closed, was floating in the leaked solution".This was observed at the hospital, prior to patient use.The device contained 4500mg fluorouracil in 0.9% sodium chloride.There was no patient involvement.No additional information is available.
|
H4: the lot was manufactured between june 26, 2023 ¿ june 27, 2023.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the provided photographic samples showed fluid inside a closed bag that contained the infusor which suggested a leak may have occurred.The reported condition was verified.Also during visual inspection, the blue winged capped was observed detached.Therefore, the cause of the leak was due to a use-error from a detached blue winged luer cap that was not securely connected after filling and priming.Per the product label (ifu, instructions for use), ¿ensure that the winged luer cap is securely connected after filling and priming.¿ a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|