DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported the cardiosave intra-aortic balloon pump (iabp) would not power on while it was plugged into wall.There was no harm reported.
|
|
Manufacturer Narrative
|
Corrected data: b5, b6, b7, d10, h6 (impact code).Updated data: b4, g3, g6, h2, h10, h11.
|
|
Event Description
|
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) would not power on while it was plugged into wall.There was no patient involved.
|
|
Manufacturer Narrative
|
The field service engineer (fse) stated that after talking with biomed, the transport unit was found to be out of the cart, not allowing it to charge the batteries.So after re-inserting the transport into the cart the batteries started charging.The unit passed all functional testing.No onsite visit was required and it was worked out over the phone.
|
|
Event Description
|
N/a.
|
|