Model Number 0998-00-0800-53
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported the cardiosave intra-aortic balloon pump (iabp) would not power on while it was plugged into wall.There was no harm reported.
 
Manufacturer Narrative
Corrected data: b5, b6, b7, d10, h6 (impact code).Updated data: b4, g3, g6, h2, h10, h11.
 
Event Description
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) would not power on while it was plugged into wall.There was no patient involved.
 
Manufacturer Narrative
The field service engineer (fse) stated that after talking with biomed, the transport unit was found to be out of the cart, not allowing it to charge the batteries.So after re-inserting the transport into the cart the batteries started charging.The unit passed all functional testing.No onsite visit was required and it was worked out over the phone.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key19099599
MDR Text Key340117764
Report Number2249723-2024-01528
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/12/2024
Supplement Dates Manufacturer Received04/16/2024
04/29/2024
Supplement Dates FDA Received04/18/2024
04/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/21/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.; UNKNOWN.
-
-