• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD SOLIS HPCA PUMP; PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD SOLIS HPCA PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 2110
Device Problems Inaccurate Flow Rate (1249); Infusion or Flow Problem (2964)
Patient Problem Insufficient Information (4580)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown; no information has been provided to date.One device was received.No physical damaged was found on the device.A review of the history log confirmed that no miss setting, or other errors related to delivery failures were identified.During functional testing, the pump accuracy was within specification however, it confirmed that it was close to the upper limit of the reference value and tended to overdose.The complaint was confirmed.Possible flow rate accuracy failure related to pump mechanism.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Adjustment of the expulsor.Performed all function tests and all passed.
 
Event Description
It was reported that the flow accuracy exceeds.There was patient involvement and unknown patient harm/adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS HPCA PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19099790
MDR Text Key340805880
Report Number3012307300-2024-02547
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0300-09JP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2024
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-