Updated field: b4, d9, g3, g6, h2, h3, h4, h6(type of investigation, investigation findings, investigation conclusions),h10.Corrected fields: d1, d4.A getinge field service engineer (fse) was dispatched to evaluate the unit.The (fse) reported that the they were able to reproduce the issue.The fse adjusted the connections of the display pcb, video receiver pcb and all connections of the main board.The unit passed all functional and safety tests and was returned to customer in good working condition.
|