DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Catalog Number 0684-00-0575 |
Device Problems
Leak/Splash (1354); Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the intra-aortic balloon (iab) was inserted in the afternoon, and by that evening blood was seen in the helium tubing.Additionally, there were "gas gain", "gas loss", and "catheter restriction" alarms.The iab was removed without issue after 9-10 hours of use.The pa noted that there was a large kink in the off-label sheath that was used.It was stated that it was possible that this was the cause of the issue.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
Initial reporter occupation: (b)(6).Additional reporters: (b)(6).The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Reference complaint #(b)(4).
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and between the catheter and the sheath.The returned sheath was not a maquet product.The extender and pressure tubing were also returned.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal, pressure and extender tubing was performed and a leak was detected on the membrane approximately 2.3cm from the rear seal measuring 0.025cm in length.The reported problems was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint (b)(4).
|
|
Event Description
|
N/a.
|
|