BIOTRONIK AG, BUELACH, SWITZERLAND PULSAR-18 5/170/135; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 366836
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2024
Event Type  malfunction  
Manufacturer Narrative
Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
 
Event Description
The pulsar-18 self-expandable stent system was selected for treatment of a severe calcified lesion (95 percent stenosis degree) in the moderately tortuous ostial left sfa.After pre-dilatation with two passeo-18 balloons, the pulsar-18 was introduced, but the device could not pass the lesion.The affected device was withdrawn and scrapped at the hospital.The procedure was finished with another stent system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSAR-18 5/170/135
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key19104836
MDR Text Key340901306
Report Number1028232-2024-02047
Device Sequence Number1
Product Code NIP
UDI-Device Identifier07640130415102
UDI-Public(01)07640130415102(17)2609
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P160025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number366836
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07233721
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-