MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL EU 4.5X37MM STENT 12 MM DW TIP; INTRACRANIAL NEUROVASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL EU 4.5X37MM STENT 12 MM DW TIP; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENC453712
Device Problems Premature Activation (1484); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2024
Event Type  malfunction  
Manufacturer Narrative
Product complaint#: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section e1.Initial reporter phone: (b)(6).Section h3 - the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2024-00393.
 
Event Description
As reported by the field, during a endovascular embolization, an eu 4.5x37mm stent 12 mm dw tip intracranial stent (enc453712, 8302881) became impeded in the prowler select plus 150/5cm (606s255x, 31122635) and could not pass through the microcatheter (mc).The stent was released automatically, and the stent body was found to be separated prematurely from the delivery wire.The physician removed the microcatheter and stent from the patient and switched new devices to complete the surgery.Additional event information received on 08-apr-2024 indicated that they were able to torque the device.The resistance was felt at the hub.Other devices were successfully used with the concomitant device prior to the encountered resistance.There were no procedural delays to the event.
 
Manufacturer Narrative
Product complaint # (b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3, h6 and h10.Complaint conclusion: as reported by the field, during a endovascular embolization, an eu 4.5x37mm stent 12 mm dw tip intracranial stent (enc453712, 8302881) became impeded in the prowler select plus 150/5cm (606s255x, 31122635) and could not pass through the microcatheter (mc).The stent was released automatically, and the stent body was found to be separated prematurely from the delivery wire.The physician removed the microcatheter and stent from the patient and switched new devices to complete the surgery.Additional event information received on 08-apr-2024 indicated that they were able to torque the device.The resistance was felt at the hub.Other devices were successfully used with the concomitant device prior to the encountered resistance.There were no procedural delays to the event.A non-sterile eu 4.5x37mm stent 12 mm dw tip was received contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and it was noted that the stent was already detached from the unit and this together with the introducer component were not returned for evaluation.The delivery wire was in good condition (i.E., no kinks, bents, or elongations).The delivery wire was subjected to dimensional analysis and all measurements were found to be within specification, including those specifications that control the attachment and delivery of the stent.Therefore, device failure is not suspected to be a contributing factor.The customer complaint regarding a stent being prematurely detached was confirmed since the stent was noted as already separated from the delivery system; based on this condition, the issue regarding a stent being impeded in the microcatheter cannot be evaluated through functional testing.The stent must be inside the introducer tube to perform the functional analysis.Additionally, the returned component did not present damage that suggest that it was forcibly advanced.With the limited information available, a conclusive cause cannot be determined; however, it is possible that clinical and procedural factors, including device manipulation and operator's technique, may have contributed to the reported failure.At this time, there is no evidence to support that the issue reported in the complaint is a result of a defect inherently related to the device.The stent component might have gotten lost sometime during the post-operative handling or decontamination of the device.If additional information or components are received at a later time, this investigation will be reassessed accordingly.Lake region medical did review the device history records relative to the manufacturing, inspecting and packaging of the lot 8302881.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations: if resistance is met during manipulation, determine the cause of resistance before proceeding.Do not apply undue force if resistance is encountered at any point during stent manipulation.Withdraw the unit and advance a new one.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EU 4.5X37MM STENT 12 MM DW TIP
Type of Device
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19104878
MDR Text Key340907737
Report Number3008114965-2024-00393
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
H60001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberENC453712
Device Lot Number8302881
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PROWLER SELECT PLUS 150/5CM.
-
-