Catalog Number 5000-01-01 |
Device Problems
Failure to Pump (1502); Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that the water of the arctic sun device did not cool down.Per follow up information received via mail on 08apr2024, it was reported they repaired the device on the customer site.Defective parts were mixing pump and chiller pump.Defective parts were replaced, and the problem was resolved.
|
|
Manufacturer Narrative
|
The reported issue was confirmed.The root cause of the reported issue is a failed mixing pump.A dhr review is not required as the reported event is not an out of box failure and therefore the reported event is not manufacturing related.The reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that the water of the arctic sun device did not cool down.Per follow up information received via mail on 08apr2024, it was reported they repaired the device on the customer site.Defective parts were mixing pump and chiller pump.Defective parts were replaced, and the problem was resolved.
|
|