• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE Back to Search Results
Catalog Number 5000-01-01
Device Problems Failure to Pump (1502); Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2024
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the water of the arctic sun device did not cool down.Per follow up information received via mail on 08apr2024, it was reported they repaired the device on the customer site.Defective parts were mixing pump and chiller pump.Defective parts were replaced, and the problem was resolved.
 
Manufacturer Narrative
The reported issue was confirmed.The root cause of the reported issue is a failed mixing pump.A dhr review is not required as the reported event is not an out of box failure and therefore the reported event is not manufacturing related.The reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the water of the arctic sun device did not cool down.Per follow up information received via mail on 08apr2024, it was reported they repaired the device on the customer site.Defective parts were mixing pump and chiller pump.Defective parts were replaced, and the problem was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19106409
MDR Text Key340905269
Report Number1018233-2024-02075
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5000-01-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2024
Initial Date FDA Received04/15/2024
Supplement Dates Manufacturer Received05/02/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-