MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN STAT; ARCTIC SUN DEVICE Back to Search Results
Catalog Number 60000000
Device Problem Lack of Effect (4065)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Event Description
It was reported that targeted temperature (tt) was 36c, but patient had been at 36.7c for a while.They wanted to make sure they were not missing anything.Water temperature (wt) was 12.9c, flow rate was 2.3lpm.Patient needed medium pads, but they only had large, they were overlapped in the back, low water limit 4c, mixing pump command was 25%.Alarm 104 (manual control end) and 112 (confirm return to cooling phase) in event log.Patient arrived 35c and was not to 36c after several hours, so they put the device in manual control with a water target of 40c.When the patient reached 36c they changed back to automatic mode.Explained that if this comes up again in the future, they did not need to place the device in manual.If the target was higher than the patient's current temperature the water temperature would increase on its own to increase patient temperature.Patient was unstable, and prone.They were unable to check pad coverage on the front.The patient was too unstable to give tylenol.No signs of shivering or seizure, they would start eeg tomorrow.Device responding appropriately suggested treating heat generation per facility protocol.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN STAT
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19106434
MDR Text Key340905687
Report Number1018233-2024-02065
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741161513
UDI-Public(01)00801741161513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number60000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-