The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation, where visual inspection, electrical screening, irrigation and microscopic tests were conducted in accordance with bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax in addition, the shaft was observed bent.The device was connected to the carto 3 system and error 105 appeared on the system due to an open circuit in the tip area.Then, an electrical test was performed, and current leakage was observed due to corrosion residues on the internal printed circuit board (pcb) connector.Finally, an irrigation test was performed, and no water leakage was detected.A manufacturing record evaluation was performed for the finished device number 31195442l, and no internal action was found during the review.The reddish material inside the pebax could be related to the force issue reported by the customer.In addition, the reported electrical issues were also confirmed due to the corrosion found therefore, customer complaint was confirmed.The potential cause of the damage on the pebax could be related to the usage of the device during procedure; however, this cannot be conclusively determined.The potential cause of the corrosion cannot be determined.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: pc-(b)(4).
|
It was reported that the patient underwent an atrial fibrillation ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, there was electrocardiogram (ecg) signal interference, a force issue, and current leakage.The signal interference( noise )was observed on the carto 3 system only, and the medical team was still able to monitor the patient's heart rhythm through other devices (body surface, anesthesia monitor, defibrillator).Additionally, the force value could not be zeroed and current leakage was detected on the patient interface unit (piu) rl input.The catheter was inside of the patient's body when these issues occurred.A second catheter was used to complete the operation.There was no adverse event reported on patient.
|