Catalog Number 254400521 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/10/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the distal femoral locking jig is snapped at the red connecting piece.It was unknown, if there was a surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received.A.Was there any adverse consequences that affected the patient, because of the reported event? no.B.Was there a surgical delay? if yes, what is the duration of the delay? no delay.
|
|
Manufacturer Narrative
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # = > (b)(4).Investigation summary = > it was reported that the distal femoral locking jig is snapped at the red connecting piece.It was unknown, if there was a surgical delay.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found that attune distal femoral jig was broken from one portion of the slide tube assy, the broken fragment was not returned for evaluation.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the attune distal femoral jig would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: d4 (lot, catalog, udi), h4 corrected: h3.
|
|