Catalog Number EX061203C
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video and photos were provided for review.The investigation of the reported event is currently underway.H10: d4: (expiration date: 04/2025).H11: section a: through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately one month post a stent placement, the stent was allegedly found to be broken.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified in d2 and g4.H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation.X-ray images and videos were provided for both tied complaints demonstrating patient treatment.Single stent struts are not visible due to poor image quality so that a stent fracture cannot be confirmed.Although single stent struts are not visible, the contour of the stent in proximal position confirms stent twisting.Based on the investigation of the provided information, the investigation is closed as confirmed for stent twisting.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instruction for use closely describe holding and handling of the system for regular deployment, in particular the instruction for use state: 'confirm that the introducer sheath is secure and will not move during deployment.Firmly hold the black system stability sheath throughout deployment.Do not hold the silver stent delivery sheath.Do not constrict the stent delivery sheath during stent deployment'.In regard to pta the instruction for use state: 'predilation of the lesion should be performed using standard techniques' and 'post stent expansion with a pta catheter is recommended'.In regard to access and accessories the instruction for use state: 'gain femoral access utilizing a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035 inch (0.89 mm) diameter guidewire'.Stent fracture was found mentioned as a potential adverse event that may occur.H10: (expiration date: 04/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Event Description
It was reported that approximately one month post a stent placement, the stent was allegedly found to be broken.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified.Manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation.X-ray images and videos were provided demonstrating patient treatment.Single stent struts are not visible due to poor image quality so that a stent fracture cannot be confirmed.Although single stent struts are not visible, the contour of the stent in proximal position confirms stent twisting.An 8f introducer with 0.035" guidewire were used for access, the vessel was not tortuous/calcified, and the lesion was pre and post dilated.The system was correctly held at the stability sheath, the user did not experience difficulty during deployment, and there was no irregularity observed in the stent strut structure; torque was neither felt nor exerted on the system.Based on the investigation of the provided information, the investigation is closed as confirmed for stent twisting.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use closely describe holding and handling of the system for regular deployment, in particular the instructions for use state: 'confirm that the introducer sheath is secure and will not move during deployment.Firmly hold the black system stability sheath throughout deployment.Do not hold the silver stent delivery sheath.Do not constrict the stent delivery sheath during stent deployment'.In regard to pta, the instructions for use state: 'predilation of the lesion should be performed using standard techniques' and 'post stent expansion with a pta catheter is recommended'.In regard to access and accessories the instructions for use state: 'gain femoral access utilizing a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035 inch (0.89 mm) diameter guidewire'.Stent fracture was found mentioned as a potential adverse event that may occur.Expiration date: 04/2025.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately one month post a stent placement in the superficial femoral artery, the stent was allegedly found to be broken.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT XL VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19113232
MDR Text Key340799225
Report Number9681442-2024-00094
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEX061203C
Device Lot NumberANHR0890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-