Model Number 4543
Device Problem Under-Sensing (1661)
Patient Problem Syncope/Fainting (4411)
Event Date 01/01/2023
Event Type  Injury  
Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event Description
It was reported that this patient with this right atrial (ra) lead and newly implanted unknown device presented to the emergency department unresponsive initially with a history of having unknown device issues three days prior.This patient was discharged and then today was unresponsive in their bedroom.Upon review, this patient had runs of ventricular tachycardia (vt) on the external monitor and chest compressions were almost initiated.Some episodes with sustained rates above and below the detection zones and did not meet the criteria for therapies.Intermittent atrial undersensed beats were observed.This patient was transferred to the specialist that implanted the device.Due to limited information, it is unknown if this device and ra lead remain in service and further follow up is not possible at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
OSCOR INC
3816 desoto blvd
palm harbor FL 34683
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19115475
MDR Text Key340270396
Report Number2124215-2024-22875
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number4543
Device Catalogue Number4543
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/29/2024
Initial Date FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization;
-
-