SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problems Material Separation (1562); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the following were found during the evaluation; due to wear of grip, water tightness is lost; there is a gap between acoustic lens and ultrasound probe; and the light guide bundle has breakage.The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was observed during the device inspection, that the ultrasound gastrovideoscope had a gap of adhesive on the distal end / stain entered inside the lens through the gap.There is a gap between acoustic lens and ultrasound probe.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and because the phenomenon was not duplicated during device evaluation, the root cause of the phenomenon could not be identified.As result of confirming instruction manual, it was found that cautions were described about phenomenon.Instructions evis exera ii ultrasound gastrovideoscope olympus gf type uct180.Important information - please read before use ¿ "do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient".¿ "do not attempt to bend the endoscope¿s insertion section with excessive force.Otherwise, the insertion section may be damaged".¿ "do not apply shock to the distal end of the insertion section, particularly the ultrasound transducer and the objective lens surface at the distal end.Visual abnormalities may result".¿ "do not twist or bend the bending section with your hands.Equipment damage may result".¿ "do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19117747
MDR Text Key341260234
Report Number3002808148-2024-03539
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356346
UDI-Public04953170356346
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-