It was reported that the temperature deviation alarm was occurring due to the patient temperature being more than 0.5 degrees from target temperature.It was reported that the device was in auto mode, with a rectal temperature of 35.4 and is correlating to a secondary temperature.The wraps are cool to the touch and the water temperature is 40 degrees.It was reported that they were originally cooling the patient and met the target temperature and now were re-warming however, the patient was not warming to the goal as fast as expected.The patient was not adversely affected and no clinically relevant delay in treatment was reported.
|