• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION NON VENTED HIGH VOL.INLET,N/S; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION NON VENTED HIGH VOL.INLET,N/S; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938173
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was moisture contamination in an unspecified quantity of non-vented high-volume inlets.The moisture contamination was discovered in the iv room prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Eight (8) devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NON VENTED HIGH VOL.INLET,N/S
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
PLASTICOS Y MATERIAS PRIMAS PYMPSA
3609-2 juan de la barrera
parque industrial el alamo
guadalajara, jalisco 44490
MX   44490
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19119255
MDR Text Key340305267
Report Number1416980-2024-01730
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475783
UDI-Public(01)00085412475783
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938173
Device Lot Number803681
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-