An adverse skin reaction was reported with wear of the abbott diabetes care (adc) device.The customer experienced symptoms described as "burning sense, non-sleeping, pain on the moving arm and the touching arm hurts" at the sensor site and had phone contact with a healthcare professional (hcp) who prescribed flucloxacillin for treatment.There was no report of death or permanent impairment associated with this event.
|
The most probable root causes associated with this incident are the adhesive, including the adhesive irritating the user¿s skin, or misuse, including improper site selection and repeatedly using the same application site to place the sensor.These conditions are mitigated through the freestyle product labelling.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.The dhrs (device history review) for the libre sensor and sensor kit were reviewed and the dhrs showed the libre sensor and sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
An adverse skin reaction was reported with wear of the abbott diabetes care (adc) device.The customer experienced symptoms described as "burning sense, non-sleeping, pain on the moving arm and the touching arm hurts" at the sensor site and had phone contact with a healthcare professional (hcp) who prescribed flucloxacillin for treatment.There was no report of death or permanent impairment associated with this event.
|