BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that three (3) exactamix 500 ml eva tpn bags were leaking from the administration port.This occurred during setup.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: three (3) actual devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The bags underwent functional leak testing and after several minutes the bags were filled with tap water leaks were observed originating from the spike port bonding area in all three bags.The reported condition was verified in all three bags.The cause of the condition could not be determined; however, the most probable cause was due to inadequate, or lack of cyclohexanone applied to the spike port cap tube when it was inserted into the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19121923
MDR Text Key340447823
Report Number1416980-2024-01749
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938738
Device Lot Number60426384
Was Device Available for Evaluation? No
Date Manufacturer Received03/22/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-