Model Number 102096-81A
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2024
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On march 17, 2024, senseonics was made aware of an incident where the user received an early sensor replacement alert resulting in an early sensor removal.
 
Manufacturer Narrative
The sensor replacement alert was presented to the user on march 17, 2024, on day 45 after insertion.The returned sensor was tested in-house, and it revealed a loss of chemical performance which caused a decline in the signal modulation.The system correctly disabled the sensor due to performance deviation, and the system's self-test functions were working normally.The root cause of the failure was due to the sensor's hydrogel oxidation.As part of resolution, the rma was authorized for sensor replacement.B4.Date of this report 17 may 2024.G3.Date received by the manufacturer? 15 may 2024.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key19122089
MDR Text Key340447871
Report Number3009862700-2024-00607
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491022950
UDI-Public817491022950
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/13/2024
Device Model Number102096-81A
Device Catalogue NumberFG-7200-00-301
Device Lot NumberWP09278
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2024
Date Manufacturer Received03/17/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-