BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G125
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 03/23/2024
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) delivered an inappropriate shock.Boston scientific technical services (ts) was consulted and upon further review of the episode, discussed it looked like external electromagnetic interference (emi) that was oversensed by the device and inappropriately treated.No adverse patient effects were reported.At this time, this device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19122306
MDR Text Key340753071
Report Number2124215-2024-23117
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589126
UDI-Public00802526589126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/26/2023
Device Model NumberG125
Device Catalogue NumberG125
Device Lot Number149770
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2024
Initial Date FDA Received04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-