Investigation findings: items returned: - n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: based on a review of the last 2 years of complaint data, and at the time of this investigation, no systemic issues have been identified for this batch number that would have caused or contributed to the reported event.Ifu review (additional information can be found in the ifu): potential complications: potential complications include but are not limited to the following: vessel puncture site hematoma, aneurysm perforation or rupture, hemorrhage, edema, thromboemboli, transient ischemic attack, ischemic stroke, neurologic deficits, parent artery occlusion, ischemia, vessel dissection or perforation, vascular thrombosis, vasospasm, device migration or misplacement, premature detachment, headache, post-embolization syndrome, infection and death.Warnings: use of the web embolization device in anatomy with severe tortuosity, stenosis, or vessel narrowing may result in difficulty or inability to deploy the subject web embolization device and can lead to damage of the web embolization device or microcatheter.The web embolization device must be delivered only through a compatible microcatheter with a ptfe inner surface coating.If an incompatible microcatheter is used, damage to the web embolization device and delivery device may occur and necessitate removal of both the web embolization device and microcatheter from the patient.Advance and retract the web embolization device slowly.Do not advance the delivery device with excessive force.Determine the cause of any unusual resistance.Remove the web embolization device if excessive friction is noted and check for damage.Do not rotate the delivery device during or after delivery of the web embolization device.Rotating the web embolization device may result in damage or premature detachment.Procedure - instructions for use: introduction and deployment of the web embolization device.18.Open the rhv on the microcatheter to accept the introducer sheath.19.Insert the introducer sheath through the rhv.Seat the distal tip of the introducer sheath at the distal end of the microcatheter hub and close the rhv lightly around the introducer to secure the rhv to the introducer.Do not over-tighten the rhv.Only insert the introducer into the microcatheter hub until a slight resistance is felt.Do not over insert.20.Push the web embolization device into the lumen of the microcatheter.Use caution to avoid catching the web embolization device on the junction between the introducer sheath and hub of the microcatheter.21.Push the web embolization device through the microcatheter until the proximal end of the delivery device meets the proximal end of the introducer sheath.22.Loosen the rhv.23.Retract the introducer sheath just out of the rhv.24.Close the rhv around the delivery device.25.Slide the introducer sheath completely off the delivery device using care not to kink or damage the delivery system.26.Carefully advance the web embolization device until the distal end of the web embolization device reaches the last marker on the microcatheter.27.Reposition the tip of the microcatheter so that it sits just at the neck of the aneurysm.Do not insert the microcatheter completely inside the aneurysm.28.Under fluoroscopic guidance, slowly advance the web embolization device out the tip of the microcatheter.Continue to advance the web embolization device into the lesion until optimal deployment is achieved.The following may require repositioning or removal of the web embolization device and/or repositioning of the microcatheter.A.If the web embolization device size is not appropriate, remove and replace with another web embolization device.B.If undesirable movement of the web embolization device is noted following placement and prior to detachment, remove the web embolization device.Movement of the web embolization device may indicate that the web embolization device could migrate once it is detached.C.If the web embolization device (implant) does not fully open: i.Retrieve the implant, reposition the microcatheter further proximally and re-deploy the implant to allow more room for expansion, or ii.Replace the implant with another implant of the same or alternate size.The web embolization device should not be retracted and deployed more than twice.After two attempts, remove the web embolization device and replace with another web embolization device of the same or alternate size.To minimize the potential risk of thrombus, do not allow an inappropriately sized or non-optimally positioned web embolization device to reside in the aneurysm significantly beyond the activated clotting time (act).Experience has shown that thrombus can also prevent the web from full deployment and recapture.To minimize the risks of potential complications, the status of the patient¿s anti-platelet medication regimen should be considered when making a decision to remove the entire system from the aneurysm for replacement by a new web embolization device.29.Angiographic assessment should always be performed prior to detachment to ensure that the web embolization device is not markedly protruding into the parent vessel.30.The web embolization device should be placed with the proximal surface (center of proximal marker) aligned with the aneurysm neck and the proximal marker extending beyond the neck.31.If redeploying, retract web embolization device so that there is no contact with the aneurysm before attempting redeployment.32.Tighten the rhv to prevent movement of the web embolization device.33.Verify that the distal portion of the delivery device is not under tension or compression prior to detachment.This could cause the microcatheter tip to move resulting in aneurysm or vessel rupture.Investigation conclusion: the physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.H3 other text : device location is unknown, per the clinical trial protocol, the devices are not tracked and are thus not available for return and investigation.
|