Catalog Number 05942861001 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/24/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The serial number for meter a is (b)(6).The serial number for meter b is (b)(6).The product was returned for investigation.On a regular basis accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable glucose results from 2 accu-chek inform ii meters compared to laboratory results using an unknown method.The result from meter a (uu14478370) was 62 mg/dl at 15:46.The laboratory result was 117 mg/dl at 15:45.The result from the patient's dexcom device was 121 mg/dl at 15:45.The result from meter a uu14478370 was 133 mg/dl at 18:15.The result from meter b uu14508915 was 159 mg/dl at 18:15.The result from the patient's dexcom device was 251 mg/dl at 18:15.The result from the patient's "personal meter" was 212 mg/dl at 18:15.The type of meter was not provided.The laboratory result was 270 mg/dl at 18:33.The result from meter a (uu14478370) was 190 mg/dl at 18:42.The result from the patient's dexcom device was 250 mg/dl at 18:42.
|
|
Manufacturer Narrative
|
Testing was performed with the returned test strips using glycolyzed blood.All results were acceptable.All testing with the returned strips produced acceptable results and no strip defects were observed.The investigation did not identify a product problem.
|
|
Search Alerts/Recalls
|