• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problems Failure to Interrogate (1332); Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  Injury  
Event Description
It was reported that the patient assisted their wife during an mri (magnetic resonance imaging).After the mri, the patient received an alert that their implantable cardioverter defibrillator (icd) disconnected from the mobile application.The patient was brought into clinic and it was noted the icd was in backup mode.The device was explanted and replaced.The patient condition was stable.
 
Event Description
New information received indicates that therapies were deactivated due to the back-up mode.An attempt to reset and restore the device was attempted but it was unsuccessful.The patient was stable.
 
Manufacturer Narrative
The reported event of backup operation was confirmed.Final analysis of device image found the device went into backup mode due to a power-on-reset as a result of off-label mri exposure.After the device was restored from backup mode, functional testing was performed.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19127239
MDR Text Key340408698
Report Number2017865-2024-38947
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public(01)05415067032010(10)S000087896(17)240831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberS000087896
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/17/2024
Supplement Dates Manufacturer Received04/23/2024
05/03/2024
Supplement Dates FDA Received04/24/2024
05/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1458Q/86 QUARTET LEADS; 2088TC/52 TENDRIL LEADS; 7122Q/65 DURATA TACHY MRI LEADS
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexMale
-
-