DATASCOPE CORP. - FAIRFIELD LINEAR 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0475
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2023
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
 
Event Description
N/a.
 
Manufacturer Narrative
Initial reporter: (b)(6) , patient safety specialist / (b)(6), patient safety lead, (b)(6) event site name: (b)(6) hospital - rochester the device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.Reference complaint # (b)(4).
 
Event Description
Reported via medwatch #(b)(4) received 25-march-2024: registered nurse (rn) monitoring patient with right femoral balloon pump in place.Machine alarmed that it was in standby, as well as for loss of gas.Rn assessed balloon tubing and found blood specks to be throughout the tubing.Clamped tubing immediately and notified provider of rupture.Patient had to undergo expedite procedure.The iab was in use for one day.There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key19130959
MDR Text Key340858106
Report Number2248146-2024-00227
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567106564
UDI-Public10607567106564
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0684-00-0475
Device Lot Number3000313067
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARDIOSAVE / (B)(6)
-
-