ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72081-01
Device Problem High Readings (2459)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 04/03/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care (adc) has been submitted.
 
Event Description
A high readings issue was reported with the adc device.Customer reported initially being asymptomatic and falling asleep; however, after receiving unspecified sensor scan result perceived to be high compared to readings (unspecified) on another adc meter; the customer experienced loss of consciousness and was unable to self-treat.The customer self-presented themselves at hospital but prior to that obtained a sensor scan result reading of approximately 2.5 mmol/l.At the hospital a healthcare professional (hcp) obtained unspecified readings on their hcp meter and it was reported they treated the customer with five (5) injections of glucose.No further treatment or medication was reported.There was no report of death or permanent impairment associated with this event.A sensor reading of 8.70 mmol/l was compared against a reading of 4.00 mmol/l on another adc meter and plotted on a parkes error grid which fell into the "c" zone showing the difference in values to be clinically significant.It is unknown when these readings were taken in relation to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19131235
MDR Text Key340449002
Report Number2954323-2024-13168
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K213996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number72081-01
Was Device Available for Evaluation? No
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-