Model Number 71992-01
Device Problem Low Readings (2460)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406)
Event Date 03/18/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A low readings issue with the abbott diabetes care (adc) device was reported.The customer received unspecified low sensor scan result in comparison to unspecified readings obtained on unspecified source.The customer experienced loss of consciousness or seizure and received sugar water and 6 units of unspecified medication administered via pumps by a non-healthcare professional for treatment.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key19131559
MDR Text Key340446938
Report Number2954323-2024-13163
Device Sequence Number1
Product Code QLG
UDI-Device Identifier00357599800000
UDI-Public(01)00357599800000(17)2024-06-30(10)KTP007868(91)71992-01-0006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number71992-01
Was Device Available for Evaluation? No
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-